Current version
EZSTATSG3.6
|
|
Previous versions
EZSTATSG3.5
EZSTATSG3.4 (1/2022)
Download template from Supplementary Data
Download template from Supplementary Data
- All features from EZSTATSG3.4 plus:
- Bug fix regarding ISTD in Waters importation.
- Support for SCIEX OS data.
- Calibration data from Thermo TraceFinder, Agilent MassHunter, Waters UNIFI, and SCIEX MultiQuant and SCIEX OS reports can be directly imported.
- Reports must be in *.xlsx, *.xls, *.csv, or *.txt format.
- For Thermo data:
- Reports must be in Calibration Report format.
- Thermo QC data can be imported if it is in the default QC Report format. Files must be formatted as follows:
- Each trial must have it's own folder named "Trial_1", "Trial_2"... OR "Trial 1", "Trial 2"... (not case sensitive).
- Each individual file name must contain the QC level i.e., "High", "Low", "Mid" (not case sensitive).
- For SCIEX MultiQuant and SCIEX OS data:
- There must be an ISTD Area column.
EZSTATSG3.4 (1/2022)
- Calibration data from Thermo TraceFinder, Agilent MassHunter, Waters UNIFI, and SCIEX MultiQuant calibration reports can be directly imported.
- Reports must be in *.xlsx, *.xls, *.csv, or *.txt format.
- Target compound no longer needs to be in the same tab index or column across workbooks.
- For Thermo data:
- Reports must be in Calibration Report format.
- Thermo QC data can be imported if it is in the default QC Report format. Files must be formatted as follows:
- Each trial must have it's own folder named "Trial_1", "Trial_2"... OR "Trial 1", "Trial 2"... (not case sensitive).
- Each individual file name must contain the QC level i.e., "High", "Low", "Mid" (not case sensitive).
- For SCIEX MultiQuant data:
- There must be an ISTD Area column.
- Bug fix for < 7 calibrators using Agilent importation.
- Bug fix on Ion Suppression page.
- Decimals supported to accommodate Agilent data.
- Updated color scheme and formatting for readability.
- Feature addition - if the c term of the quadratic equation is in error, each term will be flagged red
- Bug fix - Nonnumeric values do not result in an error in userform
- Bug fix - User is prompted to change analytes/assays containing "/" or "\"
Download template from Supplementary Data
- Flexible five-, six-, or seven-point calibration curves
- Six weighted quadratic calibration models
- Standardized residuals by use of frequency plots overlaid with the normal distribution function along with five-number summary data
- Processed sample stability function
- Implementation of Visual Basic for Applications (VBA) in Excel UserForms
Download template from Supplementary Data
- Weighted linear calibration models
- Bias and precision data for the lower limit of quantitation
- Quality controls
- Dilution integrity
- Ion suppression